EasyManuals Logo
Home>Keithley>Test Equipment>S530

Keithley S530 User Manual

Keithley S530
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
S530 Parametric Test System Test Subroutine Library User's Manual Section 3: Test subroutine library reference
S530-907-01 Rev. A / September 2015 3-55
res
This subroutine calculates the resistance of a two-terminal resistor (force I, measure V).
Usage
double res(int hi, int lo, int sub, double itest)
hi
Input
lo
Input
sub
Input
itest
Input
Returns
Output
The calculated resistance:
Details
This subroutine calculates the resistance of a two-terminal resistor by forcing a current and
measuring the voltage. The voltage is limited to 30 V.
If a positive substrate pin is specified, the substrate is grounded. If a positive substrate pin is not
specified, the substrate is left floating.
A delay is incorporated into the res subroutine; this delay is the calculated time required for stable
forcing of itest with a 30 V voltage limit.
Source-measure units (SMUs)
SMU1: Forces itest, 30 V voltage limit, measures voltage
Example
result = res(hi, lo, sub, itest)
Schematic

Other manuals for Keithley S530