VIII. TECHNICAL SPECIFICATIONS
30.0 TECHNICAL DATA
The system complies with the requirements of IEC 60601-1-2, 4
th
edition, regarding electromagnetic compatibility. Sur-
rounding equipment must follow IEC 60601-1-2.
CAUTION: Mobile telephones and other radiating equipment can interfere with the function of the system and
can therefore cause safety hazards.
30.1 Irradiation
Table 30.1.1
Irradiaon
Test Item Standard Test Performed Power (Voltage/Fre-
quency)
Remarks
Emissions
Radiated RF emis-
sions Test
CISPR11 Class A Yes
480VAC/60Hz
120VAC/60Hz
Conducted emis-
sions Test
CISPR11 Class A Yes
30.2 Positioner Movement
Table 30.2.1
Posioner Movement
Test Item Standard Test Performed
Power (Voltage/Fre-
quency)
Remarks
Immunity
Radiated RF elec-
tromagnetic eld
immunity test
IEC61000-4-3 Yes
480VAC/60Hz
120VAC/60Hz
Electrical fast
transient/Burst
Immunity Test
IEC61000-4-4 Yes
Power Frequency
Magnetic eld
immunity test
IEC61000-4-8 Yes
Immunity Test
to Conducted
disturbances In-
duced by RF eld
IEC 61000-4-6 Yes
Surge Immunity
Test
IEC 61000-4-5 Yes
Voltage dips,
short interrup-
tions, and voltage
variations immu-
nity Test
IEC 61000-4-11 Yes
120VAC/60Hz

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta KDR AU DDR and is the answer not in the manual?

Konica Minolta KDR AU DDR Specifications

General IconGeneral
BrandKonica Minolta
ModelKDR AU DDR
CategoryMedical Equipment
LanguageEnglish

Related product manuals