132
KMHA 500-000099
Revision D
KDR
TM
AU-DDR Advanced U-Arm System with
Dynamic Digital Radiography (KDR AU DDR)
VIII. TECHNICAL SPECIFICATIONS
Table 30.5.1
Radiated Immunity Test Frequencies
Equipment used Test Frequency Test Level
Teseq CBA Amplier 870 MHz 28 V/m
Teseq CBA Amplier 930 MHz 28 V/m
Amplier research 15S1G3 Am-
plier
1720 MHz 28 V/m
NOTE: The emissions characteristics of this device make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this
device may not offer adequate protection to radio-frequency communication services. The user may need to
take mitigation measures, such as relocating or re-orienting the equipment.
NOTE: Based on studies of the CPI generator, EMC is assumed to be the same or similar for the Sedecal genera-
tor.
WARNING! This device has not been tested for radiated RF immunity over the entire frequency range of 80
MHz to 6000 MHz. This device has been tested for radiated RF immunity only at selected frequencies (see Table
30.5.1.) Use of nearby emitters at other frequencies may result in improper operation.
30.6 Environmental Requirements
Table 30.6.1
Environmental Requirements
Storage
Temperature
Storage
Humidity
Storage
Pressure
Operating Tem-
perature
Operating Humid-
ity
10-40
o
C
35%-75% 500-1060 hPa
10-30
o
C
35%-70%