135
KMHA 500-000099
Revision D
KDR
TM
AU-DDR Advanced U-Arm System with
Dynamic Digital Radiography (KDR AU DDR)
VIII. TECHNICAL SPECIFICATIONS
31.4 List of Exposure Parameters (Reference)
The list of exposure values in Table 31.4.1 is for reference only. The presentation of an image depends on the requirements
of the physician. Exposure parameters must be adjusted for optimal results based on body habitus, exposure geometrics,
etc.
Table 31.4.1
List of Exposure Parameters
Area
17x17 HD/FNB 17x17 HQ/KDR
Tube
Voltage
(kVp)
mAs
Value
(mAs)
SID
(cm)
Grid
Ratio
Tube
Voltage
(kVp)
mAs
Value
(mAs)
SID
(cm)
Grid
Ratio
Skull AP 80 11 120 8:1 80 14 120 8:1
Chest PA 120 1.6 180 12:1 120 2 200 12:1
Chest LAT 125 7 200 12:1 125 9 200 12:1
Abdominal AP
(upright)
85 8 120 8:1 85 10 120 8:1
Cervical Spine
AP
70 6 120 8:1 70 7 120 8:1
Cervical Spine
LAT
70 6 120 8:1 70 7 120 8:1
Lumbar Spine
AP
75 12 120 8:1 75 16 120 8:1
Shoulder Joint
AP/Axial
75 5 120 8:1 75 6 120 8:1
Ribs 70 5 120 8:1 70 6 120 8:1
Humerus AP 60 3 120 - 60 4 120 -
Pelvis AP 85 8 120 8:1 85 10 120 8:1
Hip Joint Axial 75 15 120 8:1 75 20 120 8:1
Femur AP 70 6 120 8:1 70 8 120 8:1
Knee AP 55 6 120 - 55 8 120 -
NOTE: SID (Source Image Distance) affects the source to skin distance. Changing the source to skin distance will
change the radiation dose to the patient. A shorter distance with the same loading factors will increase the dose
to the patient.
NOTE: When using dynamic mode, a lower target ms range of 4-8 ms for the Sedecal Generator, and 4-10 ms for
the CPI generator, is required. 4-6 ms is the ideal range.
NOTE (Sedecal generators only): If an AEC technique is used and the exposure values or the positioning
of the patient are incorrect, then an extremely short exposure, less than 1ms, will result. The reported post
parameters will be not accurate. This can be recognized by the post parameters reporting the back up
timer is milliseconds (ms), while no back up timer error is displayed.
This only has inuence on the generator post parameters and the not EI or DAP values which will display
correctly to the exposure.
Example: Direct exposure of one of the selected AEC chambers with a beam energy setting above 100kV
or an AEC kV/ma exposure setting signicantly higher than advised or recognized by international stan-
dards can result in exposures less than one millisecond.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta KDR AU DDR and is the answer not in the manual?

Konica Minolta KDR AU DDR Specifications

General IconGeneral
BrandKonica Minolta
ModelKDR AU DDR
CategoryMedical Equipment
LanguageEnglish

Related product manuals