VIII. TECHNICAL SPECIFICATIONS
30.3 Standby Mode
Table 30.3.1
Posioner Movement
Test Item Standard Test Performed Power
(Voltage/Frequency)
Remarks
Immunity
Electrostatic Dis-
charge Immunity
Test
IEC 61000-4-2 Yes 480 VAC/60Hz
Electrostatic Dis-
charge Immunity
Test
30.4 Power Input Information
Table 30.4.1
Power Input Informaon
Component (Item) Power Frequency
Generator 480VAC±10% 60Hz
Positioner 120VAC±10% 60Hz
PSU(Power Supply Unit) 120VAC±10% 60Hz
Ultra Console (PC) 120VAC±10% 60Hz
30.5 Radiated Immunity Test Frequencies
Table 30.5.1
Radiated Immunity Test Frequencies
Equipment used Test Frequency Test Level
Cobra WXST CB Radio 27.12 Mhz 10 v/m
Yaesu Transceiver 52.66 MHz 10 v/m
Yaesu Transceiver 145.83 MHz 10 v/m
Yaesu Transceiver 121.72 MHz 10 v/m
Yaesu Transceiver 430.0 MHz 10 v/m
Midland GXT FRS 462.50 MHz 10 v/m
Teseq CBA Amplier 40.68 MHz 10 v/m
Teseq CBA Amplier 433.92 MHz 10 v/m
Teseq CBA Amplier 915 MHz 10 v/m
Amplier research 15S1G3 Am-
plier
2.45 GHz 28 v/m
Table 9 Test Frequencies
Teseq CBA Amplier 385 MHz 27 V/m
Teseq CBA Amplier 450 MHz 28 V/m
Teseq CBA Amplier 710 MHz 9 V/m
Teseq CBA Amplier 745 MHz 9 V/m
Teseq CBA Amplier 780 MHz 9 V/m
Teseq CBA Amplier 810 MHz 28 V/m

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta KDR AU DDR and is the answer not in the manual?

Konica Minolta KDR AU DDR Specifications

General IconGeneral
BrandKonica Minolta
ModelKDR AU DDR
CategoryMedical Equipment
LanguageEnglish

Related product manuals