5.3 Electromagnetic compatibility
Air mattress system is intended for hospitals except for near active HF surgical equipment and the RF shielded room of a medical
system for magnetic resonance imaging, where the intensity of EM disturbances is high.
Air mattress system has dened no essential performance.
WARNING!
It is recommended to avoid the use of this device next to or in block with other device, because it could lead to 
improper operation. If such use is needed, this device and the other equipment should be under surveillance
to verify proper operation. (Does not apply for compatible medical bed from LINET)
List of used cables:
1. Mains cable, maximum length 6 m
WARNING!
Use of the accessories, converters and other cables, than specied and provided by manufacturer of this air 
mattress system could lead to increase of electromagnetic emission or lower the electromagnetic immunity of 
this air mattress system and lead to improper operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables and external antenna) should 
not be used closer than 30 cm (12 inches) from any part of this air mattress system CliniCare 100 HF, including
cables specied by manufacturer. Otherwise this could lead to deterioration of functionality of this air mattress 
system.
WARNING!
Do not overload the air mattress system (Maximum Load), respect the duty cycle (INT.) and consider chapter 11
Maintenance in order to maintain the basic safety with regard to electromagnetic disturbances for the expected 
service life.
5.3.1 Manufacturer instructions - electromagnetic emissions
Emission Test Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations / icker emissions
IEC 61000-3-3
Complies

Table of Contents

Other manuals for LINET CliniCare 100 HF

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET CliniCare 100 HF and is the answer not in the manual?

LINET CliniCare 100 HF Specifications

General IconGeneral
BrandLINET
ModelCliniCare 100 HF
CategoryMedical Equipment
LanguageEnglish

Related product manuals