11 Maintenance
WARNING!
Risk of injury when working on the mattress replacement system!
Ensure that the mattress replacement system is disconnected from the mains power prior to installation, putting
into service, maintenance and deinstallation.
WARNING!
Risk of injury due to defective mattress replacement system!
Have a defective mattress replacement system repaired immediately.
If the defect cannot be repaired, do not use the mattress replacement system.
CAUTION!
Material damage due to incorrect maintenance!
Ensure that maintenance is performed exclusively by manufacturer´s customer service or by authorised service
personnel certied by the manufacturer.
If the defect cannot be repaired, do not use the mattress replacement system.
11.1 Regular maintenance
Perform regularly visual check of the product (with delivery note if necessary).
Ask service department of the manufacturer for addition of the original spare parts if some product parts are missing.
Ask service department of the manufacturer for replacement of any damaged product parts by the original spare parts.
Check inside and outside of mattress and outside of SCU for mechanical damage and signs of severe wear and tear.
Check if mattress and SCU are working properly.
Check external air lter in side of SCU for dust and dirt. If dust or dirt is visible, replace lter.
11.2 Spare Parts
The serialt label is located on the SCU and on the mattress. The serial labels contain information for claims and ordering replace-
ment parts.
Information about spare parts is available from:
Manufacturer´s customer service
Sales department
11.3 Safety Technical Checks
WARNING!
Risk of injury due to incorrect safety technical checks!
Ensure that safety technical checks are performed exclusively by manufacturer´s customer service or by authori-
sed service personnel certied by the manufacturer.
Ensure that the safety technical checks are recorded in the service and maintenance log.
Safety technical check of the mattress replacement system must be performed at least once every 12 months.
The procedure for performing the safety technical check is stipulated in EN 62353:2014.
NOTE On request, the manufacturer will provide service documentation (e.g. circuit diagrams, component part lists, descriptions,
calibration instructions etc.) for service personnel for the repair of ME equipment designated by the manufacturer as repairable by
service personnel.

Table of Contents

Other manuals for LINET CliniCare 100 HF

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET CliniCare 100 HF and is the answer not in the manual?

LINET CliniCare 100 HF Specifications

General IconGeneral
BrandLINET
ModelCliniCare 100 HF
CategoryMedical Equipment
LanguageEnglish

Related product manuals