9.5 CPR – Cardipulmonary resuscitation
Fig. CPR Function
Before starting CPR with SCU connected:
► Press red CPR button.
► Remove air connector plug from SCU.
► Mattress starts to deate and resuscitation procedure can commence.
9.6 Power failure
WARNING!
Risk of injury due to power failure!
► Seek clinical advice immediately as alternating pressure therapy is not possible during power failures.
In case of power failure the mattress will return to reactive only operation. If the air connectors and SCU are left attached the
mattress will remain inated for at least 12 hours. No active modes are available without power.
9.7 System faults
System faults are indicated by the amber light on the SCU.
NOTE: During initial ination the low pressure indication will come on until the mattress has achieved its minimum
pressure. This does not mean there is a fault unless the indicator has been on for more than 20 minutes after a power cycle.
CliniCare 100 HF SCU
Meaning Indicator
Low pressure Low pressure indicator is illuminated
Power failure Power switch is not illuminated