D9U003VC0-0101_0536
10.3.3 Cleaning the air pipe
► Wipe air pipe cover and exposed pipes with cleaning agent or disinfectant.
► Let air pipe dry.
-or-
► Remove the air pipe cover (see chapter Removing the Mattress Cover) and clean it as stated above (for the top mattress
cover see chapter Machine washing of CliniCare 100HF removable top mattress cover), if full disinfection is required.
► Clean the pipes in general accordance with chapter Cleaning Top and/or Base Cover.
► Let air pipe dry.
10.3.4 Cleaning the SCU
► Remove lter.
► Before cleaning of SCU, cover the air connector in order to prevent anything from penetrating into the air
► connector during cleaning.
► Wash the lter or replace it with the new one.
► Wipe SCU with disinfectant. Wipe SCU with cold water. It is recommended to use the cleaning wipes.
► Let SCU and lter dry.
► Reinsert lter.
10.4 Removing the Mattress Cover
► Carefully open zip under side skirt of mattress cover on foot end of mattress.
► (MCM model only. Carefully disconnect the 2 MCM air pipes at the foot end of the mattress!)
► Remove top part of mattress cover. Inspect cover and clean if necessary.
After cleaning the mattress cover:
► Reinstall mattress cover by reversing the process described above.
10.5 Removing the Air Pipe Cover
► Carefully pull the cover from the (elasticated) SCU end connector block, and draw along the pipes lengths.
► Carefully pass the 3 connector elbows through the holes to separate from the cover.
► Machine wash the cover as per chapter Machine washing of CliniCare 100HF removable top mattress cover.
After cleaning the air pipe parts:
► Reinstall air pipe cover by reversing the process described above.