3-2
NOTE
Save the packing case and packaging material as they can be used if the equipment must be reshipped.
3.1.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this manual.
The environment where the equipment is used shall be reasonably free from noises, vibration, dust, corrosive,
flammable and explosive substances. If the equipment is installed in a cabinet, sufficient space in front and behind shall
be left for convenient operation, maintenance and repair. Moreover, to maintain good ventilation, the equipment shall
be at least 2 inches (5 cm) away from around the cabinet.
When the equipment is moved from one place to another, condensation may occur as a result of temperature or
humidity difference. In this case, never start the system before the condensation disappears.
NOTE
Make sure that the operating environment of the equipment meets the specific requirements. Otherwise,
unexpected consequences, e.g. damage to the equipment, could result.
3.1.3 Installing the Battery
To install the battery:
1. Press down the release button and slide the battery door to the right as indicated to remove the battery door.
2. Align the battery pins with the battery connecter, slide the battery into the battery compartment, and press until
you hear it click into the place.
3. Cover the battery door on the compartment, and slide to the left until you hear it click into the place.
NOTE
Check the expiration date displayed on the disposable battery. Remove from use if the battery is expired.
Make sure the battery door is reinstalled properly to protect the equipment and battery.
3.1.4 Connecting Electrode Pads
1. Plug the pads connector into the pads socket.
2. Place the pads package into the pads compartment properly and carefully.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D1 and is the answer not in the manual?

Mindray BeneHeart D1 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D1
CategoryMedical Equipment
LanguageEnglish

Related product manuals