A-6
Pace Pulse
Pace pulse markers
Pace pulses meeting the following conditions are labelled with a PACE marker:
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
Pace pulse rejection
Meets the requirements of ANSI/AAMI EC13-2002: section 4.1.4.1 and 4.1.4.3.
The following pulses will be rejected.
Amplitude:
Width:
Rise time:
Minimum input slew rate:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
10V/s RTI
A.4 Power Supply Specifications
Rechargeable Battery (new and fully charged, at 20 C of ambient temperature)
Battery type
14.8V/3Ah, smart lithium ion battery, rechargeable and free of maintenance, one battery can be
configured
Charge time Less than 2.5 hours to 90% and less than 3 hours to 100% with BatteryFeed 20 charger station
Run time
Work mode Work time Testing condition
Monitoring ≥ 12 hours
LCD brightness set to low, wireless function off, not
performing defibrillation charges or discharges, and audio off
Defibrillation
≥300 discharges 200J discharges at a frequency of 3 times/min
≥200 discharges 360J discharges at a frequency of 3 times/min
Battery fuel gauge 5 LEDs indicating the current battery charge level
Remaining charge
after “Low Battery”
is reported
At least 20 minutes of ECG monitoring (under the work condition of low LCD brightness, with wireless
function turned off, not performing defibrillation charges or discharges, and audio off) and at least 10
200J discharges
Disposable Battery (new, at 20 C of ambient temperature)
Battery type Disposable, free of maintenance
Shelf life (before
insertion)
5 years from date of manufacture when stored and maintained as required
Standby life (after
insertion)
4 years when stored and maintained as required
Run time
Work mode Work time Testing condition
Monitoring ≥ 12 hours
LCD brightness set to low, wireless function off, not performing
defibrillation charges or discharges, and audio off
Defibrillation
≥300 discharges 200J discharges at a frequency of 3 times/min
≥200 discharges 360J discharges at a frequency of 3 times/min
Battery fuel gauge Battery symbol on the display indicating the current battery level
Remaining charge
after “Low Battery”
is reported
At least 20 minutes of ECG monitoring (under the work condition of low LCD brightness, with wireless
function turned off, not performing defibrillation charges or discharges, and audio off) and at least 10
200J discharges