9-3
Menu Item Options/Range Default Remark
Language
ENGLISH, SIM. CHINESE, FRENCH,
GERMAN, ITALIAN, SPANISH,
PORTUGUESE, RUSSIAN, CZECH,
DUTCH, CROATIA
/
Three languages are available on your
equipment, CHINESE, ENGLISH, and your
local language. For English or Chinese
speaking countries, the default language
options are CHINESE, ENGLISH, and FRENCH.
Bilingual Option On, Off On
Disabled when [ECG Display] is set to [On]*
Default Startup Mode* AED, Manual AED /
Voice Recording On, Off Off /
Voice Volume Auto, High, Low Auto /
Brightness Auto, Outdoor Mode, Indoor mode Auto /
9.6.2 AED Setup Menu
Menu Item Options/Range Default Remark
Shock Series 1, 2, 3 1 /
Energy 1
Adult 100, 150, 170, 200, 300, 360J 200 J
≤Energy 2
Pediatric 10, 15, 20, 30, 50, 70, 100J 50J
Energy 2
Adult Energy 1 to 360J 300 J
Cannot be less than Energy 1
Pediatric Energy 1 to 100J 70
Energy 3
Adult Energy 2 to 360J 360 J
Cannot be less than Energy 2
Pediatric Energy 2 to 100J 100J
NSA Action Monitor, CPR CPR /
Voice Prompt Interval Off, 30s, 60s, 90s, 120s, 150s, 180s 30s /
Initial CPR* On, Off Off
When set to [On], the system enters
CPR mode directly after startup.
ECG Display* On, Off On
When set to [On], the [Bilingual
Option] setting is disabled.
9.6.3 Manual Defib Setup Menu (For Pro Only)
Menu Item Options/Range Default
Default Energy for Adult 100J, 150J, 170J, 200J, 300J, 360J 200J
Default Energy for Pediatric 10J, 15J, 20J, 30J, 50J, 70J, 100J 50J
Sync After Shock Yes, No No