13-4
Source Alarm Message Alarm Level I Cause and solution
Pads Abnormal Low B The type of pads is recognized, but the one-wire
communication failed. Re-plug the pads. If the problem
persists, call for service.
Pads Expires Low B The pads have expired. Replace the pads.
Pads expiring soon Low B The pads are expiring soon. Replace the pads timely.
Load Config Err Low A An error occurred when loading configuration file. Reconfigure
the equipment. If the changes cannot be saved, call for service.
Operation Mode Error Low B When starting the main control, the obtained default startup
mode is inconsistent with that from the IO. Call for service.
Note: The special technological alarms cannot be paused or silenced, and the alarm volume is unchangeable. These
alarms stop only when the alarm condition is eliminated.