Safety Precautions 1-3
1.4 Safety Precautions
Please observe the following precautions to ensure patient and operator’s safety when
using this system.
DANGER:
DO NOT use flammable gasses, such as anesthetic gas or
hydrogen, or flammable liquids such as ethanol, near this
system, because there is danger of explosion.
WARNING:
1. Do connect the adapter power plug of this system and
power plugs of the peripherals to wall receptacles that
meet the ratings indicated on the rating nameplate.
Using a multifunctional receptacle may affect the
system protective grounding performance, and
cause the leakage current to exceed safety
requirements.
Use the cable provided with this system to connect the
printer. Other cables may result in electric shock.
You must use the power adapter provided with the
system; otherwise electric shock may result.
You can only adopt the power supply method provided
by Mindray, other power supply modes (e.g. using a
UPS) may result in electric shock.
2. Connect the protective grounding conductor before
turning ON the system. Disconnect the grounding cable
after turning OFF the system. Otherwise, electric shock
may result.
3. For the connection of power and grounding, follow the
appropriate procedures described in this operator’s
manual. Otherwise, there is risk of electric shock. DO
NOT connect the grounding cable to a gas pipe or
water pipe; otherwise, improper protective grounding
may result or a gas explosion may occur.
4. Before cleaning the system, disconnect the power cord
from the outlet. System failure and electric shock may
result.
5. This system is not water-proof designed. DO NOT use
this system in any place where water or any liquid
leakage may occur. If any water is sprayed on or into
the system, electric shock or device malfunction may
result. If water is accidentally sprayed on or into the
system, power off the system immediately and contact
Mindray Customer Service Department or sales
representative.