EasyManuals Logo
Home>Mindray>Medical Equipment>M6

Mindray M6 User Manual

Mindray M6
357 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #103 background imageLoading...
Page #103 background image
Image Optimization 5-23
5.7 PW/CW Doppler Mode Optimization
PW (Pulsed Wave Doppler) mode or CW (Continuous Wave Doppler) mode is used to
provide blood flow velocity and direction utilizing a real-time spectral display. The
horizontal axis represents time, while the vertical axis represents Doppler frequency shift.
PW mode provides a function to examine flow at one specific site for its velocity, direction
and features; while CW mode proves to be much more sensitive to high velocity flow
display. Thus, a combination of both modes will contribute to a much more accurate
analysis.
5.7.1 PW / CW Mode Exam Protocol
1. Select a high-quality image during B mode or B+ Color (Power) scanning, and adjust
to place the area of interest in the center of the image.
2. Press <PW>/<CW> to enter the sampling state,
The sampling parameters will be displayed in the image parameter area on the
left part of the screen as follows:
The sampling correction angle value will be displayed above the image, as
described in the following table:
PW Sampling Line
Adjustment
SV Size SV 3.0
Angle
Angle 0°
SVD SVD 10.5cm
CW Sampling Line
Adjustment
Angle
Angle 0°
CW Focus Depth SVD 13.2cm
3. Set the SVD by rotating the trackball; adjust the angle and SV size according to the
actual situation.
4. Press <PW>/<CW> again or <Update> to enter PW/CW Mode and perform the
examination. You can also adjust the SV size, angle and depth in the real-time
scanning.
5. Adjust the image parameters during PW/CW mode scanning to obtain optimized
images.
6. Perform other operations (e.g. measurement and calculation) if necessary.
If you choose "Enter 1D Mode Directly" in "[Setup]->[Image Preset]->[Other]", then
the sampling line will be displayed at all times in B mode images, and pressing <PW>/
<CW> will directly enter PW/ CW mode.
5.7.2 PW/CW Mode Image Parameters
In PW/ CW mode scanning, the image parameter area in the upper left corner of the
screen displays the real-time parameter values as follows: