10-8 System Diagnosis and Support
The battery status is displayed in “Preset”-“System Information”. See the table below.
Battery status Condition
Description
“GOOD” No damage to the battery or the
protection recoverable.
The fully charged battery capacity
(FCC) is larger than 40% of rated
voltage.
The battery status is good, and can
be used normally.
“WEAK” No damage to the battery or the
protection recoverable.
The fully charged battery capacity
(FCC) is less than 40% of rated
voltage.
Cycling times of the battery< 300
Discharge or charge the battery fully.
Restore some of the battery
capacity.
“RELACED” No damage to the battery or the
protection recoverable.
The fully charged battery capacity
(FCC) is less than 40% of rated
voltage.
Cycling times of the battery > 300
Battery capacity goes down after
long-time use. It affects the battery
performance and endurance time.
Replace the battery as necessary.
connect AC power supply, or Power-off”"
System will Power-off in 60s "
error, Battery supply”
occurs when the
power is supplied by
AC. Check the battery
connection or replace
the battery.
“Battery over-temperature, Please
connect AC power supply, or Power-off”"
System will Power-off in 60s "
*** “Left/Right Battery
over-temperature, Current
left/right
temperature:?,Battery
Manufacture Access value:
left:0x????,right: 0x????"
Use the adaptor to
supply the power or
shut down the device
to make the battery
cooler.
“Battery error! Battery cannot be used!
***“Left/Right Battery error!
Battery Manufacture Access
value: left:0x????,right:
0x????”
Right or left battery is
damaged or protected.
It can be recognized
via Manufacture
Access value. If it is
under protection, it
would restore after a
time period. If it is
damaged, it needs to
replace the battery.
N/A
*** “Battery cycle is more than
300, Current cycle: left:??,
right:??”
The usage time of
batteries is too long. It
is recommended to
replace the battery.