System Diagnosis and Support 10-7
Step Procedures Increment Description
Possible
reason for
starting
stagnation
Initializing
ultrasoun
d
peripheral
Configure file dialog box
Initialize print library
Battery monitoring, PHV monitoring
Load print task icon
Configure low consumption
Configure USB manager and burn manager
Initialize video review device
In
increment
of 1.
Configure
the
external.
Be careful
of the
battery.
ultrasoun
d image
Create ECG receiving thread
Set virtual machine; initialize virtual machine
Create front-back object tree (ultrasys, etc)
increments
of 2.
Pod data
Initializing
ultrasoun
d
applicatio
n
Add function package of measurement menu
Register application interface for the patient
to be measured
In
increment
of 1.
Patient
information
library
/
Initializati
on
completed
Configure the architecture of patient
information management (UPatientApp)
In
increment
of 1
/
/ Enter Doppler interface
In
increment
of 1
Doppler
interface
/
10.3 Alarming and Abnormal Information
The system has also alarming function, when the machine fails, it would pop up the alarming
dialogue box, and simultaneously generate LOG file saved in the system log, which is saved in D
disk: \M9\Log.
10.3.1 Power Error
Description: The asterisk “***” represents the time in LOG record. The format is:
2011-6-12 14:15:15
“Warning! Battery operation error! The
battery may be damaged”
*** “Battery hot unplug during
operation.”
Avoid hot swapping of
battery.
“Battery communication error! Battery
may not be used or battery capacity may
not be displayed correctly”
*** “Left/Right Battery I2C
error”
The battery error
occurs when the
power is supplied by
AC. Check the battery
connection or replace
the battery.
“Battery communication error, Please
***“Left/Right Battery I2C