Product Principle 4-7
Implements the switch of 192 array probe within the probe. Probe board only supports
128
th
channel.
4.5 PHV Power Supply Board
PHV power supply board
LDO
Flyback
circuit
Descending
circuit
Ascending
circuit
Socket
(board
to
board)
Main
board
MCU
P12V
LC
filtering
LC
filtering
LC
filtering
12VIN
12VIN
11V5VCC
CW_VIN1
CW_VIN2
CW_PWR_P
CW_PWR_N
AP100V
AN100V
PHV1P
PHV1N
PHV2P
PHV2N
UART_TX
UART_RX
UART_TX
UART_RX
CW_PWR_P
CW_PWR_N
AP100V
AN100V
PHV1P
PHV1N
PHV2P
PHV2N
DAC_CW
Over
current
and
over
voltage
signal
Over
current
and over
voltage
signal
Sampling
voltage
DAC_SDI
Power
supply
communication
&control
Figure 4-6 Diagram of PHV power supply board
Function description:
PHV power supply board is the board that provides the power to the transmitting circuit.
AP100V, AN100V, PHV1P, PHV1N, PHV2P, PHV2N, CW_PWR_P and CW_PWR_N are
all outputs.
PHV output goes into the power supply by producing the circuit via LDO.
PHV input goes into the main board after being filtered.
4.6 ECG Unit
ECG module is designed for monitoring ECG signals and displaying ECG waveforms, which
serves as reference of ultrasound images. This module can synchronize with 2D images and
color flow images in real-time display. With R wave detected, ECG signals, after being

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M9T and is the answer not in the manual?

Mindray M9T Specifications

General IconGeneral
BrandMindray
ModelM9T
CategoryMedical Equipment
LanguageEnglish

Related product manuals