4-4 Product Principle
Function description:
Circuit Unit Function
The control for
Starting and
shutting down the
device
Takes charge of starting and shutting down the device and
controls the indicator for starting and shutting down the
device
System monitoring Takes charge of battery management, fan control, voltage
detection, battery indicator control,
voltage detection and temperature detection
Wireless net
adapter
WIFI and Bluetooth
Provides two external USB 3.0 ports
SATA port SSD, mSATA type
HDMI Provides an HDMI port and supports 1920*1080 resolutions
(HD) output.
Audio Encodes and decodes the audio. Increases the power.
Signal process
FPGA
Connects CPU module with the ultrasound circuit
LCD port Connects to LCD monitor screen, displays parameter board
screen and detects of environment brightness.
Other circuits
related to
ultrasound functions
Transmitting/receiving control
Probe control
4D, TEE probe control (reserved function, to be
released)
PHV control
ECG control