SUMMARY OF FUNCTION
3. Timer input function
This function is used when you intend to take in data automatically from measuring
equipment in the same interval.
Press the PRINTER ON/OFF key, and at this state pushing the PRINTER ON/OFF
key enters the unit to this function, and the following pressed key can set the interval
time. When you finish this function, press the PRINTER ON/OFF key, while pressing
the ‘CLkey.
NOTE
1. CL’ ‘CE’ ‘ STATkeys fulfill their functions even if taking in data by the
interval timer.
2. When the interval timer is finished, if data is stored in the buffer, that data may
be printed.
3. During data input by the interval timer, if you intend to change the interval
time, finish this mode once, clear data and reset.
Each key and interval times are as follows.
Key Interval time
STAT 0.25 sec
TOL.LIMIT 1 sec
CE 5 sec
CL 30 sec
DATA 1 sec
FEED 30 minute
PRINTER ON/OFF 60 minute
NOTE
Setting 0.25 sec, 1 sec, only statistical calculation results can be printed.
Measured data can not be printed. Also, when using 0.25 sec., the data buzzer
will not sound.