83
4. Press the Up Key to display yes and press the Mode Key.
Set Value LED Display Model Basic Model
5. The warning message sure will be displayed if field calibration is performed
for the first time. OMRON does not guarantee the results of field calibration.
This warning message will not be displayed the next time field calibration is
performed on the K3NX.
Set Value LED Display Model Basic Model
6. Press the Mode Key to display the inĆt input range setting.
Set Value LED Display Model Basic Model
7. Press the Shift Key so that the K3NX will be ready for the input range.
Set Value LED Display Model Basic Model
8. Press the Up Key and input the range to be calibrated.
Set Value LED Display Model Basic Model
9. The teaching indicator will be lit and l.ref will be displayed for reference in-
put signal L.
Set Value LED Display Model Basic Model
10. Press the RESET/TEACH Key and input 1.0000 V as reference input signal
L. The teaching indicator will flash.
Set Value LED Display Model Basic Model
Maintenance Mode
Section 5-3
AUDIN - 7 bis rue de Tinqueux - 51100 Reims - France - Tel : 03.26.04.20.21 - Fax : 03.26.04.28.20 - Web : http: www.audin.fr - Email : info@audin.fr

Table of Contents

Other manuals for Omron K3NX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron K3NX and is the answer not in the manual?

Omron K3NX Specifications

General IconGeneral
BrandOmron
ModelK3NX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals