11 400-12355-000 Rev. 03
KR2i / KMPi TFF Systems
Specifications
b. Input Frequency: 47-63 Hz
c. Input Current: ~0.4A
KONDUiT Power Requirements
a. Voltage: 24 VDC
b. Current: 0.625 A
Environment
Temperature, Operating: 2°C to 50°C
Temperature, Storage: -25°C to 65°C
Chemical Resistance: Powder-coated aluminum, urethane
Pressure Range: Rated for pressure up to 75 psi (5 bar)
Readability
Conductivity
Range: 0.1 to 100 mS/cm
Accuracy: From 0.1 to 2 mS/cm ± 0.1 mS/cm; 2 to 50 mS/cm ±5% of reading; 50 to 100 mS/cm
typically ±5% of reading
Temperature
Range: 0-70°C
Accuracy: Better than ±0.2°C (typically better than 0.1°C)
Optional: UV
Output signal: 4-20 mA sourcing with 400 ohm maximum at 24 VDC; scaled to 0-2 AU with
repeatability of 1% of full scale (0.02 AU)
Typical Response Time: 1 second
Maximum Zero Shift: <2% of full scale (<0.040 AU)
Long Term Output Drift: <5% per month of full scale (<0.100 AU)
Compliance: Conforms to ANSI/UL Std 61010-1
Certified to CAN/CSA Std C22.2 No. 61010-1
This product has been tested to the
requirements of CAN/CSA-C22.2 No. 61010-1
second edition, including Amendment 1, or

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN Spectrum KR2i and is the answer not in the manual?

REPLIGEN Spectrum KR2i Specifications

General IconGeneral
BrandREPLIGEN
ModelSpectrum KR2i
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals