47 400-12355-000 Rev. 03
KR2i / KMPi TFF Systems
System Functionality
8.2.7 Ctrl Mode
Ctrl Mode is used when no Aux Pump 1 is available and additional buffer or sample is pulled into the
sealed process reservoir by vacuum. Please refer to section 8.2.2 how to setup the reservoir to pull
the vacuum. This is called "Control Mode" because the user can tell the KF Comm software when the
concentration ends and diafiltration begins instead of being done in auto mode. Please see section
below for further explanation.
In Control Mode, the user can manually tell the pump when their flowpath is routed
to run in a Concentration or Diafiltration configuration. The Data Collection and
calculations on KF Comm will reflect which of the two functions is running until
either the “Permeate Weight Stop” set-point is reached or the user presses “Stop”
on the Pump or Pump Control. Control Mode can only be accessed properly using KF
Comm on a computer.
KMPi Pump
Permeate Reservoir
Diafiltration Buffer Reservoir
Feed Scale
(Vacuum generated in sealed vessel)
Permeate Pump
(Auxiliary Pump)
Permeate Scale
Feed Reservoir
Module
Control Mode
Starting Dialogue
appears when “Start
is pressed on Pump
Control when Pump
Mode is set to
“Cntrl Mode”.
After Control Mode starts,
the user can switch between
the two functions to update
KF Comm’s calculations to
suit the selected function.
Control Mode will stop when
either “Permeate Weight
Stop” set-point is reached, or
user surpasses “Stop” on
Pump or Pump Control.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN Spectrum KR2i and is the answer not in the manual?

REPLIGEN Spectrum KR2i Specifications

General IconGeneral
BrandREPLIGEN
ModelSpectrum KR2i
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals