49 400-12355-000 Rev. 03
KR2i / KMPi TFF Systems
Pump Head Setup and Operation
a. Manual Mode/Before Running Pump Mode:
i. First Line: Current Feed Scale weight (preceded by “F:”)
ii. Second Line: Current Permeate Scale weight (preceded by “P:”)
NOTE: If KONDUiT is present and set-points are Conductivity or UV set-points, then depending on
what KONDUiT sensors are being used the lines will display current KONDUiT readings:
Line 1 - Conductivity 1
Line 2 - Conductivity 2
Line 3 - Temperature (i.e. Temperature 1 if connected. Tempeature 2 if only the Cond/Temp 2
input is used)
Line 4 - UV 1
b. After Running Pump Mode:
i. First Line: Current
starting value
ii. Second Line: Target
set-point value
iii. Fourth Line: Current function (e.g. Conc 2, Dia 2)
Pump Head Setup and Operation
9.1 Introduction
The Easy-Load and High-Performance Pump Heads are designed to be used with the TFF systems as
a simple peristaltic pump system. The Pump Heads accept different tubing sizes for a wide range of
flow rates, and the unique designs and automatic tubing retention allow for quick tubing changes.
9.2 Specifications
KR2i Pump Heads
Part Number Roller and Bearing Material Number of Rollers
MasterFlex® L/S® Tubing
ACR2-H3I-01N Stainless Steel 3 L/S® 13, 14, 16, 25, 17, 18
Typical Flow, Pressure and Vacuum Data3 roller pumps
MasterFlex®
L/S®
Tubing
Flow rate* Discharge Pressure*
Vacuum*
@ 600 rpm
in (mm) Hg
Suction lift*
@ 600 rpm
ft (m) H
2
0
@ 1 rpm
mL/Rev
@ 600 rpm
mL/min
Continuous
psig (bar)
Intermittent
psig (bar)
L/S® 13 0.06 36 25 (1.7) 40 (2.7) 26 (660) 29 (8.8)
L/S® 14 0.22 130 25 (1.7) 40 (2.7) 26 (660) 29 (8.8)
L/S® 16 0.80 480 25 (1.7) 40 (2.7) 26 (660) 29 (8.8)
L/S® 25 1.7 1000 20 (1.4) 35 (2.4) 26 (660) 29 (8.8)
L/S® 17 2.8 1700 15 (1.0) 20 (1.4) 20 (510) 22 (6.7)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN Spectrum KR2i and is the answer not in the manual?

REPLIGEN Spectrum KR2i Specifications

General IconGeneral
BrandREPLIGEN
ModelSpectrum KR2i
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals