51 400-12355-000 Rev. 03
KR2i / KMPi TFF Systems
Pump Head Setup and Operation
Pump
tubing size
mL per rev
Flow rates in LPM (GPM) at indicated rpm Maximum System
pressure psi (bar)
100
540
1 to 650
I/P 26 6.2 0.6
(0.2)
3.3
(0.9)
0.01 to 4.0
(0.002 to 1.1)
40 (2.7)
I/P 73 12.3 1.2
(0.3)
6.6
(1.7)
0.01 to 8.0
(0.002 to 2.1)
I/P 82 20.0 2.0
(0.5)
10.8
(2.9)
0.02 to 13.0
(0.005 to 3.5)40 (2.7)
20 (1.4)
I/P® 26, I/P 73 I/P® 82
Maximum continuous discharge pressure: 25 psi (1.7 bar)* 10 psi (0.7 bar)*
Maximum intermittent discharge pressure: 40 psi (2.7 bar)* 20 psi (1.4 bar)
Maximum vacuum: 660 mmHg (26 inHg)* 660 mmHg (24 inHg)*
Ma Maximum suction lift: 8.8 m (29 ft) H20* 8.2 m (27 ft) H20*
Occlusion: Adjustable (with Knob on top of Pump Head)
Maximum pump speed: 650 rpm
Normal torque load-Starting: up to 13 kg-cm (180 oz-in)
Normal torque load-Running: up to 9 kg-cm (120 oz-in)
Housing materials: Polysulfone (PSF) or Polyphenylene sulfide (PPS),
Polyester, Aluminum, Nylon.
Operating temperature‡: 0 to 40°C (32 to 104°F)
* As tested with NORPRENE®, PHARMED®, and TYGON® tubing. Values will be less with silicone and
C-FLEX.
‡ Use in this temperature range for continuous duty operation with no decrease in performance or
life. Pump heads will work outside this range with some possible reductions in performance or
product life.
Part Number
Roller and Bearing Material
Number of Rollers
MasterFlex® I/P® Tubing
ACM3-PHP-01N Stainless Steel 3 I/P® 70, 88, 89
Typical Flow, Pressure and Vacuum Data3 roller pumps
I/P Tubing
Flow rate _____________Discharge Pressure*
Vacuum*
@ 600 rpm
in (mm) Hg
Suction lift*
@ 600 rpm
ft (m) H
2
0
@ 20 rpm
L/min
@ 600 rpm
mL/min
Continuous
psig (bar)
Intermittent
psig (bar)
I/P 70
0.06
8
25 (1.7)
40 (2.7)
26 (660)
29 (8.8)
I/P 88 .40 13 20 (1.4) 35 (2.4) 26 (660) 29 (8.8)
I/P 89 .52 17 15 (1.0) 20 (1.4) 24 (610) 27 (8.3)
Occlusion: Fixed
Maximum pump speed (rpm): 650
Nominal torque load-running: up to 600 oz-in (43 kg-cm)
Housing materials: Stainless Steel (SS), polyester (PE) body and
tubing bed, polypropylene (PP) knob
Operating Temperature‡: 32°F to 104°F (0°C to 40°C)
Storage Temperature: -40°F to 149°F (-40°C to 65°C)
Humidity: 5% to 95% (non-condensing)
Dimensions (W x H x D):

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN Spectrum KR2i and is the answer not in the manual?

REPLIGEN Spectrum KR2i Specifications

General IconGeneral
BrandREPLIGEN
ModelSpectrum KR2i
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals