RIGOL Chapter 2 Command System
2-146 MSO1000Z/DS1000Z Programming Guide
:SYSTem:LOCKed
:SYSTem:LOCKed <bool>
:SYSTem:LOCKed?
Enable or disable the keyboard lock function, or query the status of the keyboard lock
function.
The query returns 1 or 0.
:SYSTem:PON
:SYSTem:PON <pon>
:SYSTem:PON?
Set or query the system configuration to be recalled when the oscilloscope is powered on
again after power-off.
The query returns LAT or DEF.
:SYSTem:PON LATest /*Set the system configuration to be recalled when the
oscilloscope is powered on again after power-off to last*/
:SYSTem:PON? /*The query returns LAT*/
:SYSTem:OPTion:INSTall
:SYSTem:OPTion:INSTall <license>
To install an option, you need to order the option to obtain the corresponding key.
Then, use the key to acquire the option license by following the steps below.
Log in RIGOL official website (www.rigol.com); click SERVICE Software
License Register to enter the product license register interface.
Input the correct key, serial number (press Utility System System Info
to get the instrument serial number) and identifying code in the product license
register interface; click Generate to acquire the option license.
<license> is the option license (the hyphens should be omitted). It is a 28-byte string
and can only include uppercase English characters and numbers.
:SYSTem:OPTion:INSTall PDUY9N9QTS9PQSWPLAETRD3UJHYA