EasyManuals Logo
Home>Rigol>Test Equipment>DS1104Z Plus

Rigol DS1104Z Plus Programming Guide

Rigol DS1104Z Plus
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #226 background imageLoading...
Page #226 background image
RIGOL Chapter 2 Command System
2-210 MSO1000Z/DS1000Z Programming Guide
:TRIGger:SPI
Command List:
:TRIGger:SPI:SCL
:TRIGger:SPI:SDA
:TRIGger:SPI:WHEN
:TRIGger:SPI:WIDTh
:TRIGger:SPI:DATA
:TRIGger:SPI:TIMeout
:TRIGger:SPI:SLOPe
:TRIGger:SPI:CLEVel
:TRIGger:SPI:DLEVel
:TRIGger:SPI:SLEVel
:TRIGger:SPI:MODE
:TRIGger:SPI:CS
:TRIGger:SPI:SCL
Syntax
:TRIGger:SPI:SCL <source>
:TRIGger:SPI:SCL?
Description
Set or query the channel source of SCL in SPI trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{D0|D1|D2|D3|D4|D5|D6|D7|D8|
D9|D10|D11|D12|D13|D14|D15|
CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Explanation
When one or more of channels D0 to D7 is turned on, CH4 cannot be selected; when one
or more of D8 to D15 is turned on, CH3 cannot be selected.
Return
Format
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14,
D15, CHAN1, CHAN2, CHAN3, or CHAN4.
Example
:TRIGger:SPI:SCL CHANnel1 /*Set the channel source of SCL to CH1*/
:TRIGger:SPI:SCL? /*The query returns CHAN1*/

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals