Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-185
:TRIGger:RUNT
Command List:
:TRIGger:RUNT:SOURce
:TRIGger:RUNT:POLarity
:TRIGger:RUNT:WHEN
:TRIGger:RUNT:WUPPer
:TRIGger:RUNT:WLOWer
:TRIGger:RUNT:ALEVel
:TRIGger:RUNT:BLEVel
:TRIGger:RUNT:SOURce
:TRIGger:RUNT:SOURce <source>
:TRIGger:RUNT:SOURce?
Set or query the trigger source in runt trigger.
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
The query returns CHAN1, CHAN2, CHAN3, or CHAN4.
:TRIGger:RUNT:SOURce CHANnel2 /*Set the trigger source to CH2*/
:TRIGger:RUNT:SOURce? /*The query returns CHAN2*/
:TRIGger:RUNT:POLarity
:TRIGger:RUNT:POLarity <polarity>
:TRIGger:RUNT:POLarity?
Set or query the pulse polarity in runt trigger.
POSitive: the instrument triggers on the positive runt pulse.
NEGative: the instrument triggers on the negative runt pulse.
The query returns POS or NEG.
:TRIGger:RUNT:POLarity NEGative /*Set the pulse polarity to negative*/
:TRIGger:RUNT:POLarity? /*The query returns NEG*/