Adjustment Procedure—2230 Service
c. ADJUST—A High Speed Timing (C703) for 1 time
marker per division over the center 8 divisions.
d. Set:
HORIZONTAL MODE B
A SEC/DIV
2 its
B SEC/DIV
1 fiS
e. ADJUST—B High Speed Timing (C713) for 1 time
marker per division over the center 8 divisions.
10. Adjust 5 ns Timing and Linearity
(C775 and C785)
a. Set:
CH 1 VOLTS/DIV 0.2 V
Horizontal POSITION Midrange
HORIZONTAL MODE
’ A
A SEC/DIV 0.05 fiS
XI0 Magnifier
On (knob out)
b. Select 10 ns time markers from the time-mark
generator.
c. Align the time markers with the vertical graticule
lines using the Horizontal POSITION control.
d. ADJUST—5 ns Timing (C775 and C785 alternately)
for one time marker every 2 divisions over the center 8
divisions of the magnified sweep.
e. CHECK—Time markers between the 2nd and 4th
vertical graticule lines should be aligned within 0.05 divi
sion. If not, a slight compromise between timing and
linearity should be made by readjusting the 5 ns Timing
capacitors (C775 and C785).
c. Adjust the A TRIGGER LEVEL control for a stable,
triggered display.
d. Use the Horizontal POSITION control to align the
2nd time marker with the 2nd vertical graticule line.
e. CHECK—Timing accuracy is within 2% (0.16 divi
sion at the 10th vertical graticule line), and linearity is
within 5% (0.1 division over any 2 of the center 8
divisions).
NOTE
For checking the timing accuracy of the A SEC/DiV
switch settings from 50 ms to 0.5 s, watch the time
marker tips only at the 2nd and 10th vertical grati
cule lines while adjusting the Horizontal POSITION
control.
f. Repeat parts c through e for the remaining A
SEC/DIV and time-mark generator setting combinations
shown in Table 5-7 under the "Normal (XI)" column.
g. Set:
A SEC/DIV 0.05 ns
XI0 Magnifier On (knob out)
h. Select 10 ns time markers from the time-mark
generator.
i. Use the Horizontal POSITION control to align the 1st
time marker that is 25 ns beyond the start of the sweep
with the 2nd vertical graticule line.
j. CHECK—Timing accuracy is within 3% (0.24 division
at the 10th vertical graticule line), and linearity is within 5%
(0.1 division over any 2 of the center 8 divisions). Exclude
any portion of the sweep past the 100th magnified
division.
11. Check Timing Accuracy and Linearity
a. Set:
k. Repeat parts i and j for the remaining A SEC/DIV
and time-mark generator setting combinations shown in
Table 5-7 under the "X I0 Magnified” column.
CH 1 VOLTS/DIV 0.5 V
XI0 Magnifier Off (knob in)
b. Select 50 ns time markers from the time-marker
generator.
I. Set:
HORIZONTAL MODE
A SEC/DIV
B SEC/DIV
X I0 Magnifier
B
0.1 its
0.05 fis
Off (knob in)
5-22