Specification—2465B/2467B Service
Table 1-1 (cont)
Characteristics
AUTO Mode Maximum Triggering Signal Period
A-SEC/DIV Setting
<10 ms
10 ms to 50 ms
>50 ms
AUTO LVL Mode Trigger Acquisition Time
Trigger Holdoff
Minimum
2467 B
2465 B
Variable
SLOPE Selection
Performance Requirements
At least 80 ms.
a
At least 16 times the A-SEC/DIV setting.
3
At least 800 ms.
a
Eight to 100 times the AUTO LVL Mode maximum triggering signal
period,
depending on the triggering signal period and waveform.
The greater of the A-SEC/DIV setting value or
1
us, within +33%
+ 500 ns to
-10%.
a
The greater of the A-SEC/DIV setting value or 2 us, within +33%
to -10%, except 1 ^s at 5 ns/div.
a
Increases trigger holdoff time to 10 to 25 times the minimum
holdoff.
Conforms to trigger-source waveform or ac power-source
waveform.
HORIZONTAL DEFLECTION SYSTEM
A Sweep Time Base Range
B Sweep Time Base Range
Timing Accuracy
Sweep Accuracy Unmagnified
At Accuracy With Cursors, Unmagnified
At Accuracy with Sweep Delay
Delay Accuracy, A Sweep Trigger to Start of B
Sweep
B-Sweep Accuracy and At Accuracy with Cursors on
B Sweep
500 ms/div to 5 ns/div in a 1-2-5 sequence of 25 steps. X10 MAG
extends maximum sweep rate to 500 ps/div.
50 ms/div to 5 ns/div in a 1-2-5 sequence of 22 steps. X10 MAG
extends maximum sweep rate to 500 ps/div.
+ 15°C to +35°C, A Sweep, with SEC/DIV at 100 ms/div or
faster.
±(0.7%
of time interval + 0.6% of full scale).
±(0.5%
of time interval + 0.3% of full scale).
±(0.3%
of time interval +
0.1%
of full scale + 200 ps).
±(0.3%
of delay setting + 0.6% of full scale) +0 to -25 ns.
Add ±0.3% of time interval to A-Sweep specifications.
'Performance requirement not checked in manual.