Specification—2465B/2467B Service
Table 1-2 (cont)
Characteristics Performance Requirements
TOTALIZE
Maximum Count
Display Update Rate
9999999.
Twice per second or once per event, whichever is slower.
DELAY BY EVENTS
Maximum Event Count
Minimum Time from Start Signal
to Any Delay Event
4194303.
4 ns.
LOGIC TRIGGER
Minimum Function-True Time
Minimum Function-False Time
4 ns.
4 ns.
ADDED DELAY TIME CHARACTERISTICS WITH C/T/T
Run After Delay
Accuracy
Triggerable After Delay
Accuracy
Minimum Measurable Delay Time
Display Update Rate
LSD
b
+ [0.0012 X (A SEC/DIV)] + [0.03 X (B Time/Div)
c
] + A
Trigger Level Error + 50 ns.
When the A Sweep is triggered by the Word Recognizer in
synchronous mode, add 100 ns for probe delay; in asynchronous
mode,
add 200 ns for probe delay.
For intervals within 70 ns to 10 times the A-SEC/DIV Setting.
LSD
b
+ [10 ppm x (measured interval)] + TJE + A-Trigger Level
Error + B-Trigger Level Error + 0.5 ns.
If the A and B Sweeps are triggered from different channels, add
0.5 ns for channel-to-channel mismatch.
When the A Sweep is triggered by the Word Recognizer in
synchronous mode, add 100 ns for probe delay; in asynchronous
mode,
add 200 ns for probe delay.
*£70 ns.
In Auto Resolution, twice per second or once for every sweep,
whichever is slower.
In 1 ns, 100 ps, and 10 ps resolution modes, the update rate
depends on the A SEC/DIV setting and the trigger repetition rate.
b
See Tables 1-3 and 1-4.
C
B Time/Div includes SEC/DIV, X10 MAG, and VAR.
d
This term assumes the trigger points are between the 10% and 90% points of the waveforms. Fall time is expressed as a negative
risetime.
1-16