DMM4040/4050
Users Manual
1-8
Compliance Information
This section lists the EMC (electromagnetic compliance), safety, and environmental
standards with which the instrument complies.
EMC Compliance
EC Declaration of Conformity—EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance
was demonstrated to the following specifications as listed in the Official Journal of the
European Communities:
EN 61326-1 2006, EN 61326-2 2006. EMC requirements for electrical equipment for
measurement, control, and laboratory use.
1, 2, 3
CISPR 11:2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic field immunity
4
IEC 61000-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
5
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
6, 7
EN 61000-3-2:2006. AC power line harmonic emissions.
EN 61000-3-3:1995. Voltage changes, fluctuations, and flicker.
European Contact.
Tektronix UK, Ltd.
Western Peninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic
interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test
object.
3
To ensure compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
The tolerance of the 100uADC, 10mADC ranges while subjected to the test field (3 V/m over the frequency range
of 80 MHz to 1 GHz, 1.4 GHz to 2.0 GHz, and 1V/m from 2.0 GHz to 2.7 GHz, with 80% amplitude modulation
at 1 kHz) can increase by +/- (0.06% of range) over the frequency range of 1.7 - 1.9GHz. (IEC 61000-4-3).
5
The tolerance of the 100uADC, 10mADC ranges while subjected to the injected test signal (3 V rms over the
frequency range of 150 kHz to 80 MHz, with 80% amplitude modulation at 1 kHz) can increase by +/- (0.08% of
range) at the frequency range of 20 - 50MHz. (IEC 61000-4-6).
6
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test
levels (IEC 61000-4-11).
7
Instrument rebooting may be experienced where the EUT takes approximately 18 seconds to recover from IEC
61000-4-11 transient immunity test.