DMM4040/4050
Users Manual
1-16
Overrange ............................................................... 20 % on all ranges except 1000 V dc, 1000 V ac
Diode, and 10 A ranges
Remote Interfaces
RS-232C, DTE 9-pin, 1200 to 230400 baud (RS-232C to USB cable available to connect the Meter to a PC USB port.
See Accessories)
IEEE 488.2
LAN and “Ethernet 10/100 base T with DHCP (for IP_ADDRess) option”
Warranty
Three years
Electrical Specifications
Accuracy specifications are valid for 6½ digit resolution mode after at least a 1-hour warm-up with Auto Zero enabled.
24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per
EN 61326-1:2000-11
DC Voltage Specifications
Maximum Input ....................................................... 1000 V on any range
Common Mode Rejection ....................................... 140 dB at 50 or 60 Hz ±0.1 % (1 kΩ unbalance)
Normal Mode Rejection .......................................... 60 dB for NPLC of 1 or greater with analog filter off and power line
frequency ±0.1 %
100 dB for NPLC of 1 or greater with analog filter on and power line
frequency ±0.1 %
Measurement Method ............................................. Multi-ramp A/D
A/D Linearity ........................................................... 0.0002 % of measurement +0.0001 % of range
Input Bias Current ................................................... <30 pA at 25 °C
Autozero Off Operation ........................................... Following instrument warm-up at calibration temperature ±1 °C and
less than 10 minutes, add error: 0.0002 % range additional error
+5 μV.
Analog Filter............................................................ When using the analog filter, specifications are relative to within one
hour of using the ZERO function for that range and NPLC setting.
DC Ratio ................................................................. Accuracy is +/- (Input accuracy + Reference accuracy), where Input
accuracy = DC Voltage accuracy for the HI to LO Input (in ppm of the
Input voltage), and Reference accuracy = DC Voltage accuracy for the
HI to LO (Sense) Reference (in ppm of the Reference voltage).
Settling Considerations........................................... Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes.
Input Characteristics
Resolution
Range Resolution
4½ Digits 5½ Digits 6½ Digits
Input Impedance
100 mV 100.0000 mV
10 μV 1 μV
100 nV
10 MΩ or >10 GΩ
[1]
1 V 1.000000 V
100 μV 10 μV 1 μV 10 MΩ or >10 GΩ
[1]
10 V 10.00000 V 1 mV
100 μV 10 μV 10 MΩ or >10 GΩ
[1]
100 V 100.0000 V 10 mV 1 mV
100 μV 10 MΩ ±1%
1000 V 1,000.000 V 100 mV 10 mV 1 mV
10 MΩ ±1%
[1] Inputs beyond ±14 V are clamped through 200 kΩ typical. 10 MΩ is default input impedance.

Table of Contents

Other manuals for Tektronix DMM4020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DMM4020 and is the answer not in the manual?

Tektronix DMM4020 Specifications

General IconGeneral
BrandTektronix
ModelDMM4020
CategoryMultimeter
LanguageEnglish

Related product manuals