Making Measurements
Measuring Resistance 4
4-9
Note
For best results, the filter may require zeroing during the
ohms function.
Refer to the “Range Keys” section in Chapter 3 of this manual for information on how to
adjust the measurement range.
Making a Four-wire Resistance Measurement
The Meter incorporates two methods of making a four-wire resistance measurement. The
traditional method is to use four meter leads to connect the Meter to the resistance to be
measured. The optional 2X4 Wire test leads simplifies the four-wire measurement so you
only have to plug in two test leads to the front-panel Input HI and LO connectors of the
meter.
To make a four-wire resistance measurement using four test leads:
1. Connect the test leads to the Meter’s input connectors as shown in Figure 4-2.
2. Press N.
0.217384
F1 F2 F3 F4 F5
caw031.eps
3. If not already highlighted as shown above, press the 4WIRE soft key to switch to a
four-wire measurement.
2
INPUT SENSE
HI
LO
FRONT
REAR
10 A
1V
400
mA
4
W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
600V CAT II
caw023.eps
Figure 4-2. Input Connections for 4-wire Resistance Measurements

Table of Contents

Other manuals for Tektronix DMM4020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DMM4020 and is the answer not in the manual?

Tektronix DMM4020 Specifications

General IconGeneral
BrandTektronix
ModelDMM4020
CategoryMultimeter
LanguageEnglish

Related product manuals