Model 2450
SourceMeter® Instrument User's Manual Section 3:
Using a remote interface
2450-900-01 Rev. E / August 2019 3-3
To allow many parallel connections to one instrument, stack the connectors. Each connector has two
screws on it to ensure that connections remain secure. The figure below shows a typical connection
diagram for a test system with multiple instruments.
To avoid possible mechanical damage, stack no more than three connectors on any one
instrument. To minimize interference caused by electromagnetic radiation, use only shielded
GPIB cables. Contact Keithley Instruments for shielded cables.
Figure 12: IEEE-488 connection example
Set the GPIB address
The default GPIB address is 18. You can set the address from 1 to 30 if it is unique in the system.
This address cannot conflict with an address that is assigned to another instrument or to the GPIB
controller.
GPIB controllers are usually set to 0 or 21. To be safe, do not configure any instrument to have an
address of 21.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2450 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2450 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2450
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals