EasyManuals Logo
Home>Tektronix>Laboratory Equipment>Keithley SourceMeter 2450

Tektronix Keithley SourceMeter 2450 User Manual

Tektronix Keithley SourceMeter 2450
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
Section
6: Leakage current and insulation resistance Model 2450 SourceMeter® Instrument
User's Manual
6-2 2450-900-01 Rev. E / August 2019
Equipment required
One 2450 SourceMeter
®
Instrument
Two triaxial cables
One capacitor for the leakage current application
One coaxial cable or other device for the insulation resistance application
One ethernet, GPIB, or USB cable for the TSP and SCPI remote command examples
Set up remote communications
You can run this application from the front panel or any of the supported communication interfaces for
the instrument (GPIB, USB, or ethernet).
The following figure shows the rear-panel connection locations for the remote communication
interfaces. For additional information about setting up remote communications, see
Remote
communications interfaces (on page 3-1).
Figure 27: 2450 remote interface connections
Device connections
Depending on the device under test (DUT), the current measurement is typically very small, usually
<10 nA. For any device, measuring leakage current and insulation resistance involves measuring very
small values. To get more accurate readings, connect the DUT to the 2450 rear panel with low-noise
triaxial cables.
Connect the DUT between the FORCE HI and FORCE LO terminals of the 2450.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2450 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2450 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2450
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals