Section
6: Leakage current and insulation resistance Model 2450 SourceMeter® Instrument
User's Manual
6-6 2450-900-01 Rev. E / August 2019
View the measurements on the front-panel graph
To view the leakage current measurements on the front-panel graph:
1. Press the MENU key.
2. Under Views, select Graph.
3. Select the Scale tab.
4. Under Y-Axis, next to Scale Format, select Log.
5. Under X-Axis, next to Method, select Show All Readings.
6. Select the Graph tab to view the graph.
The following figure shows the front-panel graph for this application.
Figure 31: Viewing the leakage current measurement on the front panel