SourceMeter® Instrument User's Manual Section 9: Measuring I-
V characteristics of solar cells
2450-900-01 Rev. E / August 2019 9-7
Figure 52: Example of solar cell measurements on the front-panel graph
Set up the solar cell I-V sweep using SCPI commands
This example sequence of SCPI commands generates an I-V sweep on a solar cell. You may need to
make changes so that this code will run in your programming environment.
In this example, the voltage is swept from 0 V to 0.55 V in 56 steps. The resulting solar cell current is
measured. The current and voltage measurements are stored in default buffer 1 (defbuffer1).
Send the following commands for this example application:
Command Description
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
SENS:CURR:RSEN ON
SOUR:FUNC VOLT
SOUR:VOLT:RANG 2
SOUR:VOLT:ILIM 1
SOUR:SWE:VOLT:LIN 0, 0.55, 56, 0.1
:INIT
*WAI
TRAC:DATA? 1, 56, "defbuffer1", SOUR, READ
Reset the 2450.
Set to measure current.
Set to measure with autorange enabled.
Set to use 4-wire sense mode.
Set to source voltage.
Set to the 2 V source range.
Set the current limit to 1 A.
Set to sweep voltage from 0 to 0.55 V in 56 steps
with a 0.1 s delay.
Initiate the sweep.
Wait until the sweep is finished.
Read the source and measure values from
.