19
Functional overview
This functional verification procedure helps to confirm the proper operation of the Spot
Vital Signs and options. This procedure supports the requirements of routine preventative
maintenance. It is not necessary to disassemble the Spot Vital Signs to perform this
procedure.
For the calibration procedures, see “Calibration” on page 25. If the Spot Vital Signs fails
certain functional tests or a circuit board is replaced, the device may require calibration. It
is necessary to disassemble the Spot Vital Signs for calibration.
Always perform this functional verification procedure after performing any calibration. This
procedure contains additional tests that are not included in calibration procedures.
Power on/off and system check procedure
Press the Power button to turn the device on or off. Upon each power up, all the LCD
segments in each display turn on briefly and two beeps sound. If the internal self-check is
successful, the display shows its normal functions (see page 15) and the device is ready
for operation. If the self-check fails, an error code is shown in the on the display.
To turn the unit off, press the Power button.
Note
Turning the unit off erases measurement data.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Spot Vital Signs and is the answer not in the manual?

Welch Allyn Spot Vital Signs Specifications

General IconGeneral
Pulse Rate MeasurementYes
SpO2 MeasurementYes
DisplayLCD
Power SourceRechargeable battery or AC power
Temperature MeasurementYes
ConnectivityOptional with wireless connectivity
Pulse Rate Range30-220 bpm
NIBP Diastolic Range20-195 mmHg
SpO2 Range70-100%
Temperature Range80 °F to 110 °F (26.7 °C to 43.3 °C)
Battery LifeApproximately 8 hours of continuous operation
Weight CapacityNot applicable
Height Measurement RangeNot applicable

Related product manuals