5
29
Troubleshooting
When the main printed circuit board (PCB) is replaced on a Spot Vital Signs version 1, the
print function is no longer available. The new function, Pressure Preset, raises or lowers
the blood pressure maximum inflation level for one measurement only. When replacing
the main PCB, replace the switch array as well.
The following table of conditions and error codes provides a quick reference of the
descriptions and probable causes of error codes.
To clear the error code:
Power the Spot Vital Signs off, wait five seconds, and power on. If the error code
reappears then power the Spot Vital Signs off and disconnect the battery for five minutes.
Reconnect the battery and power on. If the error code continues to reappear, call Welch
Allyn for a Service Notification Number (see “Technical assistance” on page 11).
Press the Blood Pressure Start/Stop button to reset flashing patient alarm conditions.
Error codes
Table 4. General
Code Description Corrective Action
E11 Internal safety violation Check patient, contact Technical Service.
C12 Ambient temperature out of range Adjust ambient temperature or device location.
C13 Battery failure Use wall transformer.
E0.0 - E9.9 Temperature module malfunction Contact Technical Service.
E42 Internal communications error Disconnect the battery and wait 5 minutes.
Reconnect the battery and then set the date and time.
E20 - E50 General internal malfunction Contact Technical Service.
Table 5. Blood Pressure
Code Description Corrective Action
C02 Auto-zero failure Check for air obstruction, limit patient movement.
C03 Inflation too rapid Check for kinked blood pressure cuff tubing, pressure
hose, or other air obstruction.
C04 Excessive inflation time Check for air leaks.
C05 Excessive noise Check patient condition, blood pressure cuff
placement, limit patient movement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Spot Vital Signs and is the answer not in the manual?

Welch Allyn Spot Vital Signs Specifications

General IconGeneral
Pulse Rate MeasurementYes
SpO2 MeasurementYes
DisplayLCD
Power SourceRechargeable battery or AC power
Temperature MeasurementYes
ConnectivityOptional with wireless connectivity
Pulse Rate Range30-220 bpm
NIBP Diastolic Range20-195 mmHg
SpO2 Range70-100%
Temperature Range80 °F to 110 °F (26.7 °C to 43.3 °C)
Battery LifeApproximately 8 hours of continuous operation
Weight CapacityNot applicable
Height Measurement RangeNot applicable

Related product manuals