EasyManuals Logo

Welch Allyn Spot Vital Signs Service Manual

Welch Allyn Spot Vital Signs
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
76 Repair test specifications Welch Allyn Spot Vital Signs
Mode, actuating the SpO
2
mode and applying any SpO
2
signal to the device. The
maximum limit is 120 mA maximum.
Fail safe test
Over pressure test
Verify that Spot Vital Signs can detect over pressure on the pneumatic system between
296.0 mmHg and 329 mmHg.
Over 15 mmHg
Verify that Spot Vital Signs can detect static pressure over 15 mmHg for 180 seconds.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Spot Vital Signs and is the answer not in the manual?

Welch Allyn Spot Vital Signs Specifications

General IconGeneral
Pulse Rate MeasurementYes
SpO2 MeasurementYes
DisplayLCD
Power SourceRechargeable battery or AC power
Temperature MeasurementYes
ConnectivityOptional with wireless connectivity
Pulse Rate Range30-220 bpm
NIBP Diastolic Range20-195 mmHg
SpO2 Range70-100%
Temperature Range80 °F to 110 °F (26.7 °C to 43.3 °C)
Battery LifeApproximately 8 hours of continuous operation
Weight CapacityNot applicable
Height Measurement RangeNot applicable