Specifications
Consider the accessories connected to the Spot Vital Signs as Type BF applied parts.
Patient population
The Welch Allyn Spot Vital Signs is designed for use with adult and pediatric patients.
Welch Allyn defines a pediatric patient as 29 days or more of age.
THE WELCH ALLYN SPOT VITAL SIGNS IS NOT INTENDED FOR USE ON NEONATES.
Welch Allyn defines neonates as children 28 days or less of age, born at term (37 weeks
gestation or more), otherwise up to 44 gestational weeks.
Blood pressure
Temperature
Cuff Pressure Range 0 mmHg to 300 mmHg
Cuff Inflation Factory
Default
160 mmHg
Systolic Range 60 mmHg to 250 mmHg
Diastolic Range 30 mmHg to 160 mmHg
Accuracy Blood pressure accuracy meets or exceeds SP10-1992 AAMI standards for
non-invasive blood pressure accuracy (AAMI standard: ± 5 mmHg mean error,
8 mmHg standard deviation). Blood pressure accuracy is validated for pressure
measurement using the upper arm only.
Determination Time Typical: 20 to 45 seconds
Maximum: 165 seconds
Pulse Rate Range 40 bpm to 200 bpm
Pulse Rate Accuracy ±5.0%
Overpressure Cutoff 305 mmHg -0/+15 mmHg
Accuracy Calibration accuracy:
+ 0.2° F (+ 0.1° C).
Range Maximum: 109.4° F/43.0° C
Minimum: 86.0° F/30.0° C
Determination Time Oral: approximately 4 seconds
Axillary: approximately 10 seconds
Rectal: approximately 15 seconds

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Spot Vital Signs and is the answer not in the manual?

Welch Allyn Spot Vital Signs Specifications

General IconGeneral
Pulse Rate MeasurementYes
SpO2 MeasurementYes
DisplayLCD
Power SourceRechargeable battery or AC power
Temperature MeasurementYes
ConnectivityOptional with wireless connectivity
Pulse Rate Range30-220 bpm
NIBP Diastolic Range20-195 mmHg
SpO2 Range70-100%
Temperature Range80 °F to 110 °F (26.7 °C to 43.3 °C)
Battery LifeApproximately 8 hours of continuous operation
Weight CapacityNot applicable
Height Measurement RangeNot applicable