-22-
6 Specification
6.1 Product Specification
6.1.1 Control Unit
Item Specification
Rotation speed
control
*Standard: 15005000rpm (Max.1000fps)
* High Speed: 150010000rpm (Max.2000fps)
* Possible for command control of rotation speed setting.
* Rotation speed setting/reading resolution:
15005000rpm: 1rpm
500010000rpm: 2rpm
* External sync is possible by pulse signal input via external
signal input terminal.
* External output terminal can send 12 pulse per one rotation
DM Exchange Command control of exchanging up to 3 DM positions in the DM
block
Shutter Control Command control or external line control of shutter on/off
External I/O
control
*Digital I/O: TTL Input CH, Contact Input 4CH, Open
Collector output 4CH, Contact output 4CH
*External shutter control, synchronization control of CSU
rotation speed, synchronization signal output, etc.
*Analogue input: 8 ports, 05V, 8Bit0255, Response 100ms.
*Analogue output: 4 ports, 05V, 8Bit0255, Response 10ms
Filter Wheel
control
(CSU-X1CU-F1)
Command control of filter positions
One control unit controls up to two filter wheels
Bright Field/
Camera port
control
(CSU-X1CU-B1)
Command control of light path switching (confocal/non-confocal)
Switch between 1
st
camera port and 2nd camera port
Interlock key
1.Installed
Manual shutter switch becomes valid.
External shutter control from PC through control unit becomes
possible.
2.Uninstalled
Shutter closes, motor stops and manual shutter switch becomes
invalid