-10-
(1) Warning label (1)
Warning sign to indicate this product belongs to the Class 3B laser product, and
indicates the product’s wavelength, maximum laser output power and laser class.
(2) Warning label (2)
Warning sign to instruct the operator must turn the laser off when he removes the
FC connector.
(3) Warning label(3)
Warning sign to indicate the laser aperture must be closed when the laser is on.
(4) Warning label (4) : For Direct view Port (Option)
Warning sign to instruct the operator avoid looking into the eyepiece of microscope
when exchanging the light path between confocal and non-confocal observation to
avoid laser exposure through the eyepiece.
(5) Warning label (5)
Warning sign to advise the necessity to use the laser in compliance with the class 3B
laser safety precautions. Specifically, it is necessary to avoid direct exposure of the
laser beam on the skin or eyes. Do not touch or look into the laser beam.
(6) Warning label (6)
Warning sign to indicate the operator must never insert the fingers or hand in the
filter wheel holes while operation.
(7) Laser key switch
The laser power supply used for CSU-X1 system employs a key switch for your safety.
The key switch is to turn off laser emission into the CSU-X1. For details, see the
user’s manual of the laser system you are using.
(8) Shutter
The shutter is for the incident laser beam. You may select full-open or close. For
non-confocal observation, always keep the shutter closed. Red LED light on the
operation panel of CSU-X1 will be OFF when the shutter is closed.
(9) Interlock
When you use the CSU-X1 without control unit (CSU-X1CU-F1), insert the interlock
key into the EXT CONTROL / INTERLOCK connector before turning on the unit. This
will disengage the interlock circuit, and you will be able to manually open/close the
shutter from the operation panel. If you use the control unit (CSU-X1CU-F1), insert
the interlock key into the EXT INTERFACE1 (at the back of the control unit). After
turn on the CSU-X1 unit, you will be able to open/close the shutter via operation panel
or control unit.