4-104
Screens
Pwr Ramp: Pulse (Option 006 Only)
Pwr Ramp: Pulse (Option 006 Only)
1. Amplitude This is a copy of the RF Analyzer Amplitude field. This is the amplitude to be
assumed at the selected input port.
Range RF IN/OUT: 27.9 to +41.0 dBm. AUX RF IN: 58.0 to +20.0
dBm.
See Also Screens: RF Generator/RF Analyzer, (RF Analyzer) Chapter 3:
Specifications
2. Fall Pos This field selects the time (relative to the center of the last bit) that the amplitude on
the amplitude envelope will be measured.
Range 0.0 µs to +56 µs.
3. Meas Frequency This field displays the frequency at which measurements are being made (Test Mode
only).