7-12
Messages
Monitoring For Protocol Failure And Recovery During Test.
Monitoring For Protocol Failure And Recovery During
Test.
The Agilent 8922M/S provides error reporting for protocol errors. These errors may
be due to mobile failure, base station failure or a faulty user configuration. It is wise
to query the error messages periodically. Particularly errors should be queried prior
to call setup, after call termination and after a handover.
When the error: ‘Call disconnected: PH Error: 0x009d’ (or 0x009f) is encountered,
a fatal protocol error has been detected by the Agilent 8922M/S and it has
reconfigured back to the BCCH. Testing may resume from a point at which the
mobile is camped.
When the error: ‘Call disconnected: PH Error: 0x009e’ (or 0x00a0) is encountered,
a fatal protocol error has been detected which requires cycling power on the Agilent
8922M/S.