Installation Azbil Corporation
2-54 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
(4) Cut the outside shield o at a point 5 mm (0.2 inch) from the outside sheath,
leaving the specied length from the end of the outside shield.
CAUTION
Do not cut into the inside sheath.
5 mm
Outside
shield
C-wire
5 mm
White
C-wire
Black
Inside sheath
Figure 2-50 Trimming the outside shield
(5) Peel o the inside sheath so that a 15 mm (0.59 inch) section remains.
CAUTION
Do not cut into the inside shield.
15 mm
C-wire
shield
White
C-wire
Black
15 mm
Figure 2-51 Trimming the inside sheath
(6) Cut the inside shield o at a point 5 mm (0.2 inch) from the edge.
5 mm
Inside
shield
5 mm
(0.2 inch)
C-wire
A-wire
B-wire
C-wire
sheath
Inside
sheath