RESmart GII BPAP System Y / U Series User Manual V2.1
24 / 42
13.2 Options of the Patient Menu and Corresponding
Descriptions
Option
Range
Description
Humidifier
Off, 1 5
There are five humidity levels available. As the numbering
increases, the humidity rises accordingly. Off means the
humidifier is turned off. The default setting is 2
Reslex
Off, 1 3
This feature enables the device to automatically reduce the
treatment pressure when the patient exhales, so as to make the
user more comfortable. The higher the numbering is, the more
pressure the device reduces. Off means this feature is
disabled. The default setting is Off
Ramp Time
0 - Max
Ramp
In order to increase comfort and help the patient fall asleep
easily, the pressure can increase gradually, when the Ramp
feature is enabled. The ramp time during which the initial
pressure rises to the prescribed treatment pressure can be
adjusted. As you turn the Knob to the nearest point, the
numbering increases or decreases by five minutes. The default
setting is 10 minutes. The screen displays a real-time
countdown of the remaining ramp time in minutes
Delay
On / Off
When the humidifier is on, this feature allows the airflow to
continue for about 15 minutes at a low pressure (about 2 hPa)
after you press the Knob to discontinue treatment. This
will blow off the vapor left in the humidifier to avoid any
damage to the device. When this feature is set to Off, which
means it is disabled, the airfolw stops delivering air instantly
after you press the Knob . The default setting is Off
Date
2000-01-01
2099-12-31
Setting date by adjusting this option
Time
00:00
23:59
Setting time by adjusting this option
Brightness
High / Low
Setting screen brightness by adjusting this option. The default
setting is High
Mask Type
Full Face;
Nasal;
Pillow;
Other
There are three mask types available, namely Full Face
(full-face mask), Nasal (nasal mask), and Pillow (nasal pillow
mask). The default mask type is Nasal,” but the patient can
choose other suitable masks as well. When selecting masks
other than the above three types of masks, the patient can
identify the masks as Other
iCode
iCode, iCode
QR, iCode
QR+
iCode provides access to the patient's compliance data during a
recent time period. The iCode mode displays data in sequences
of characters. The iCode QR and iCode QR+ mode display data
in two-dimensional codes
Use Time
0 50000
h
Use Time displays how long has the device been used by the
user. The use time can be erased