H60 Heated Humidifier User Manual V2.0
9 / 21
5. Cleaning
Clean the water chamber before first use of the humidifier or at least once every week. If the
humidifier has not been in use for a long time, clean the water chamber before reusing it.
WARNING!
• To avoid electrical shock, disconnect the power cord of the device before cleaning the
humidifier. DO NOT immerse the humidifier in any fluids.
CAUTIONS!
• Do not use solutions containing chlorinated lime, chlorine, or aromatic to clean the device
and its accessories. Liquid soap containing the humidifying agent or antimicrobials should not
be used in cleaning either. These solutions may harden cleaned materials or reduce their life.
• Do not clean or dry the device and its accessories when the temperature is higher than
80°C (176˚F). High temperatures could reduce product life.
5.1 Separating the Humidifier Top Cover from its Main
Body
Press the Humidifier Uncover Button to lift and open the top cover of the humidifier.
Continue to lift the top cover until it separates completely from the main body of the
humidifier, as shown in the figure below.
Humidifier Uncover Button
Fig. 5-1
5.2 Removing the Water Chamber
Pull the water chamber out of the main body of the humidifier horizontally, as shown in the
figure below.
Fig. 5-2