H60 Heated Humidifier User Manual V2.0
14 / 21
components in close contact with the patient such as the mask, headgear, and tube, so as to
prevent disinfectant residuals from damaging the skin or respiratory tract or causing
allergies.
• The device shall not be serviced or maintained while in use with a patient.
• Sterilization of this device and its components other than recommended is not permitted.
7. Service
The humidifier does not require routine servicing.
If the humidifier malfunctions, contact your home care provider immediately. Never attempt
to open the humidifier’s enclosure. If necessary, contact your local authorized dealer or BMC
Medical Co., Ltd. for technical support and documents.
8. Specifications
Size
Dimensions: 120 mm × 180 mm × 134 mm
Weight: < 1 kg
Water Capacity: 350 ml at recommended water level
Product Use, Transport and Storage
Operation Transport and Storage
Temperature: 5°C to 35°C (41˚F to 95˚F) -25°C to 70°C (-13˚F to 158˚F)
Humidity: 15% to 93% Non-condensing 15% to 93% Non-condensing
Atmospheric Pressure: 760 to 1060 hPa 760 to 1060 hPa
Power Requirements (when the heated humidifier is used with the main device.)
100 - 240 V AC, 50 / 60 Hz, 2.0 A max
Input Voltage
24 V 1.5 A
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IP22
Heater Settings
1 to 5 (95˚F to 167˚F / 35°C to 75°C)
Maximum Operating Pressure
40 hPa